Acutip Probe for Acupuncture Points, Small Joints, & General Uses.
Technical Specifications:
|
650-100 ACUTIP
|
810-250 ACUTIP
|
Output Power: |
100 mW |
250 mW |
Wavelength: |
650 nm (+/- 5 nm) |
810 nm (+/- 5 nm) |
Laser Type:
|
GaAlInP |
GaAlAs |
Beam Divergence: |
9 x 38 degrees |
9 x 38 degrees |
Polarization: |
Linear |
Linear |
Laser Classification: |
3B |
3B |
NOHD: |
38 cm |
38 cm |
Safety Eyewear Requirement: |
OD min at 650 nm |
OD min at 810 nm |
Aperature Size: |
6 mm |
6 mm |
Spot Size (Skin Contact):
|
5.8 x 2.2 mm (eliptical) |
5.8 x 2.2 mm (eliptical) |
Power Density: |
0.8 W/cm2 |
2.0 W/cm2 |
Delivers 1 Joule in: |
10 seconds |
4 seconds |
Dose Per Minute: |
6 Joules |
15 Joules |
NOTE: The Output Power and Power Density values listed above are as measured when the control unit is set to CW. The Output Power and Power Density will be reduced to 50% of the CW value when any other frequency is selected.